Hartwig Thim,
"Absence of the Relativistic Transverse Doppler Shift at Microwave Frequencies"
, in IEEE Transactions on Instrumentation and Measurement, Vol. 52, Nummer 2, IEEE, 10-2003, ISSN: 0018-9456
Original Titel:
Absence of the Relativistic Transverse Doppler Shift at Microwave Frequencies
Sprache des Titels:
Englisch
Original Kurzfassung:
An experiment is described showing that a 33-GHz microwave signal received by rotating antennas is not exhibiting the frequency shift ("transverse Doppler effect") predicted by the relativistic Doppler formula. The sensitivity of the apparatus used has been tested to be sufficient for detecting frequency shifts as small as 10e-3 Hz which corresponds to the value of (v/c)^2 = 5.10e-14 used in the transverse Doppler shift, experiment reported here. From the observed absence of the transverse Doppler shift it is consluded that either the time dilation predicted by the standard theory of special relativity does not exist or, if it does, is a phenomenon which does not depend on relative velocities but may be a function of absolute velocities in the fundamental frame of the isotropic microwave background radiation.
Sprache der Kurzfassung:
Englisch
Journal:
IEEE Transactions on Instrumentation and Measurement