Electrocardiographic predictors of atrial mechanical sensing in leadless pacemakers
Sprache des Titels:
Englisch
Original Kurzfassung:
Leadless pacemakers (LPs) capable of VDD pacing allow for atrioventricular synchrony through mechanical sensing of atrial contraction. However, mechanical sensing is less reliable and less predictable than electrical sensing. The purpose of this study was to evaluate P-wave amplitude during sinus rhythm from preoperative 12-lead electrocardiograms (ECGs) as a predictor for atrial mechanical sensing in patients undergoing VDD LP implantation.Consecutive patients undergoing VDD LP implantation were included in this 2-center prospective cohort study. ECG parameters were evaluated separately and in combination for association with the signal amplitude of atrial mechanical contraction (A4).P-wave amplitudes in ECG leads aVL and V1 can predict A4 signal amplitude in patients with VDD LP and therefore the probability of successful AV synchronous pacing.